GLUCOCARD X-mini plus INT-IVD mmol Operating Manual (83-04345) page 1 FRONT
10.1 Specifications
*
1
: Although whole blood samples are used for measurement, displayed results
are equivalent to plasma glucose levels.
Store the meter at 0 to 50°C (32 to 122°F).
The meter is shipped with the batteries loaded. So, you can start testing
immediately. However, the batteries may not satisfy the “Battery life”.
Product specifications and appearance are subject to change without notice.
Chapter 10 Technical Information
Product
Model
Test item
Sample
Sample size
Test strips
Unit of measure
Test range
Test time
Temperature
compensation
Battery
Battery life
Power consumption
Memory capacity
Clock accuracy
Operating
environment
Expected life
Dimensions
Weight
GLUCOCARD
TM
X-mini plus
GT-1960
Blood glucose level
Fresh capillary whole blood*
1
0.3 μL
GLUCOCARD
TM
X premium Test Strips
mmol/L
0.6 to 33.3 mmol/L
5 seconds from blood detection
Automatic compensation using built-in thermo sensor
3 V lithium battery (CR2032) × 2
Approx. 1,000 tests (Actual number of tests may be
less under some conditions of use.)
0.02 W (Max.)
500 test results
Within ±75 seconds/month
Temperature: 10 to 40°C (50 to 104°F)
Humidity: 20 to 80% RH (No condensation)
3 years (according to company data)
L102 × W33 × 18.5 mm (L4.0 × W1.3 × 0.7 inches)
Approx. 46 g (1.49 ounces)
REV. 2011.09
( 84-04345 )
Thank you for choosing the GLUCOCARD
TM
X-mini plus.
We have designed this compact blood glucose meter for faster and easier
testing of your blood glucose. We hope our product will help you manage your
diabetes. This manual explains how to use your new meter. Before testing,
carefully read through this manual and the package inserts that come with the
GLUCOCARD
TM
X premium Test Strips and GLUCOCARD
TM
X-METER
CONTROL. Pay particular attention to listed warnings and cautions. Please
keep this manual handy for future reference. If you have any questions, please
contact our authorized distributor(s) at the end of this manual.
NOTE
Store test strips in their original bottle to maintain quality. Do not transfer them
to other containers.
1.1 Intended Use
Your GLUCOCARD
TM
X-mini plus and GLUCOCARD
TM
X premium Test Strips
are for quantitatively measuring the glucose level in fresh capillary whole blood.
They are intended for use outside of the body (in vitro diagnostic use) at home
or in a clinical setting as an aid to monitor and control blood glucose levels. Do
not use them to diagnose diabetes. Also, do not alter treatment based on
the test results of this meter without instructions from your doctor or healthcare
professional.
1.2 Measurement Principle
Glucose in the blood reacts with the reagent in the test strip and this produces
a small electric current. The strength of this current is proportional to the
concentration of glucose in the blood. The meter measures this current and
calculates your glucose level.
1.3 What’s Included
GLUCOCARD
TM
X-mini plus
Carrying case
Operating Manual (this manual)
Simplified Manual / Information for Users
Multilet
TM
(lancets)*
Multi-Lancet Device
TM
(lancing device)*
GLUCOCARD
TM
X premium Test Strips*
Some kits do not include the above items with an asterisk (*).
2.2 What You Need for Each Test
Do not use this meter for your blood glucose testing if you are being tested
for galactose tolerance or xylose absorption. You may use this meter if you
are taking PAM (1-Methylpyridine-6-carbaldehyde oxime).
These treatments may produce higher test results than your actual
blood glucose level. Taking unnecessary measures to lower your blood
glucose level may induce severe low glucose symptoms such as coma.
Chapter 1 Introduction
For accurate test results, allow the meter to adjust to the surroundings:
Temperature: 10 to 40°C (50 to 104°F)
Humidity: 20 to 80% RH (Relative Humidity)
for 30 minutes before testing your blood glucose.
Do not store or use the meter where:
• There are sharp temperature fluctuations.
• Humidity is high enough to cause condensation (bathrooms, drying
rooms, kitchen, etc.).
• There is a strong electromagnetic field (microwave oven, cell phone,
etc.).
Do not use the meter after it has been dropped in liquid or liquids have
entered inside, even if dried afterwards.
Keep hands away from the test strip slot on the meter. A thermo sensor is
housed inside the meter to minimize detection of errors.
Do not connect the communication cable to the connection terminal
during testing. The meter may be damaged, leading to inaccurate test
results.
Do not apply blood directly to the test strip slot on the meter.
Do not share your meter with anyone to avoid the risk of infection.
2.1 Important Health-Related Information
2.3 Cautions on Using the Meter
Chapter 2 Before Tests
WARNING
CAUTION
2.4 Cautions on Using the Test Strips
2.5 Meter Parts
Use only the GLUCOCARD
TM
X premium Test Strips for testing with the
GLUCOCARD
TM
X-mini plus. Do not use other test strips as it causes
inaccurate test results.
Do not use test strips beyond their expiration date. The expiration date is
written on the test strip bottle next to “Exp.”.
For accurate test results, allow test strips to adjust to the surroundings:
Temperature: 10 to 40°C (50 to 104°F)
Humidity: 20 to 80% RH
for 20 minutes before testing your blood glucose.
Do not use the test strips if 6 months or more have passed since
opening the bottle.
The test strips are for single-use only. Do not use test strips that have
already absorbed blood or control solution.
CAUTION
Keep the meter, test strips and other items away from young children. Small
items are choking hazards.
WARNING
Connection
terminal
a. GLUCOCARD
TM
X-mini plus (meter)
b. GLUCOCARD
TM
X premium Test Strips
c. Lancing device d. Lancet
Strap hole
Display
with backlight
Test strip slot with light
Insert a test strip here before
testing.
Test strip
disposal lever
Battery holder
Button
83-04341