GLUCOCARD X-mini plus INT-IVD mmol Operating Manual (83-04345) page 1 REAR
This manual uses the following graphic icons to
show you how to operate the button.
2.6 Display
2.7 Inserting Test Strips into the Meter
Start with the following steps to test your blood glucose level.
1. With clean, dry hands, remove 1 test strip from the bottle.
Do not bend the test strip.
NOTE
Tightly close the bottle immediately after taking out a test strip.
Do not apply blood to the test strip before inserting it into the meter.
2. Insert the test strip fully into the test strip slot.
The meter turns on. Check that a full display appears as
in chapter 2.6.
After the most recent test result appears, the
blood symbol will flash on the screen.
If nothing appears on the screen, remove the test strip and insert it back into
the test strip slot.
NOTE
You can turn on the test strip light.
The test result that appears right after power turns on is the most recent
result without a control test flag or delete flag.
If the thermometer symbol appears on the screen, see “If a
Message Appears” in the “Information for Users”.
If you insert a test strip beyond the expiration date you set, the
blood symbol and expiration symbol will alternately flash on the
screen. If this occurs, see chapter 6.3.
3.4 Sampling from Your Fingertip
1. Remove the head cap.
2. Insert a new lancet into the holder until it stops.
NOTE
Use a lancet before its expiration date.
3. Twist off the protective cap on the lancet. Keep the
protective cap for disposing of the lancet after the test.
Next ...
Go to chapter 3 “Blood Sampling”. Draw blood and apply it to the test strip no
more than 5 minutes after you insert the test strip into the meter. Double beeps
begin to sound 4 minutes and 30 seconds after inserting the test strip. Thirty
seconds later, the meter turns itself off.
NOTE
In the ECO mode, the blood symbol flashes 2 minutes shorter than otherwise.
See chapter 7 “Meter Setup”.
If you cannot test your blood glucose due to a problem with your meter or
test strips, promptly contact your healthcare professional and distributor.
CAUTION
3. To flag the test result:
See step 2 in chapter 5.1.
This chapter briefly explains how to get a blood sample using the Multi-Lancet
Device
TM
and Multilet
TM
. If using some other lancing device to sample blood, see
the manual for that device.
3.1 Lancing Device Parts
3.2 Cautions on Using the Multi-Lancet Device
TM
Follow these cautions to reduce the risk of infection by pathogenic
microorganisms.
3.3 Selecting a Sampling Site
Alternative site testing (AST): This meter can test the glucose level of blood
from your fingertip, palm, forearm or upper arm. However, test results from
sites other than the fingertip may differ from a fingertip measurement. Consult
your doctor or healthcare professional before testing blood from the palm,
forearm or upper arm.
Use blood from: If you are going to test:
Chapter 3 Blood Sampling
Wash the puncture site with soap and water. Dry the site thoroughly
before sampling blood.
Do not share the same lancet or lancing device with anyone to avoid the
risk of infection.
Always use a new lancet. Lancets are for single-use only. Do not reuse a
lancet that you have already used.
CAUTION
Fingertip, palm,
forearm or
upper arm
Before meals
Two hours or more after meals
Two hours or more after exercise
Fingertip
When there is the possibility of blood glucose level
changing rapidly (e.g. after meals or exercise)
When experiencing symptoms of hypoglycemia
such as perspiration, cold sweats, a floating
sensation or trembling
When immediate testing is needed for suspected
hypoglycemia
When in poor physical condition such as having a
cold, etc.
Button Operations
Press the button once.
Press the button the specified number of
times.
Slide the button to the right.
Slide the button to the left.
Hold the button in the direction of the arrow
until the option you want appears on the
screen.
Hold the button in the direction of the arrow
for the specified duration of time.
Twice
Hold
Hold for
5 seconds
Left Right
Protective cap
Special band Band hooks Disposal-and-set lever Strap hole
Head cap Holder Puncture button Adjustment dial
Multilet
TM
(lancet) Multi-Lancet Device
TM
(lancing device)
Decimal point
Delete flag
Control test flag
HYPO flag
After-exercise flag
Displays the
blood symbol,
date and time,
etc.
Sound mute symbol
Before-meal flag
After-meal flag
Battery symbol
Clock counter
Meal amount flag
Thermometer symbol
This meter checks itself every time you insert a test strip. You do not need to
run frequent control tests. Do a control test if:
You suspect the meter or test strips are not working properly.
You dropped the meter.
You damaged the meter.
Your test results do not match how you feel.
You want to check performance of the meter and test strips before a blood
glucose test.
NOTE
Use only the control solution, GLUCOCARD
TM
X-METER CONTROL, to test
your meter and test strips
1. See steps 1 and 2 in chapter 2.7.
2. Select the control test flag.
As an alternative, you can attach the control test flag after the
test result appears on the screen. Test results with the control test flag are not
part of your result averages.
3. Make sure that the blood symbol is flashing on the screen.
4. Squeeze a drop of control solution to the tip of the test strip.
The test will not start if you apply your control solution to
the check window directly.
The test starts when the meter detects the control
solution. The meter counts down from 5 to 1 during the
test.
NOTE
Tightly close the control solution bottle.
Do not touch the test strip once the meter starts counting down.
5. Check your test result is within the acceptable range.
The acceptable range is shown on the package insert that come with the
control solution. If within range, the meter and the test strip are working
properly. If out of range, test the control solution again.
Chapter 8 Control Test
Do not use control solutions beyond their expiration date. The expiration
date is written on the control solution bottles next to “Exp.”.
Leave the meter and test strips at
10 to 40°C (50 to 104°F) and 20 to 80% RH,
and the control solution at
22 to 28°C (72 to 82°F)
for at least 30 minutes before running the control test.
Do not drink control solution. It is not for human consumption.
Keep control solution away from skin and eyes. Contact may cause
inflammation.
CAUTION
6. Slide the test strip disposal lever to eject the test strip.
The meter turns itself off.
If you still have test results that are out of the acceptable range, stop using
your meter, test strips and control solutions, and contact your distributor.
CAUTION
9.1 Storing Your Meter
After use, tightly close the caps of the test strip bottle and control solution bottle
to maintain their quality.
Store your meter, test strips, control solutions and manuals in your carrying
case and store it in a dry location. The correct storage temperature is 0 to 50°C
(32 to 122°F) for the meter and 1 to 30°C (34 to 86°F) for the test strips and
control solutions. Do not freeze. Avoid heat, humidity and direct sunlight.
Chapter 9 Meter Care
9.2 Cleaning Your Meter and Lancing Device
Your meter and lancing device do not need special cleaning. If your meter and
lancing device get dirty, wipe them with a soft piece of cloth moistened with
water. To disinfect these items after cleaning, wipe with a soft cloth moistened
with 70% ethanol, 70% isopropanol or diluted household bleach (0.05% sodium
hypochlorite solution).
9.3 Changing the Batteries
When the battery symbol appears on the screen, the batteries
are getting low. Before using your meter, change the batteries.
Past results remain in the memory even when the batteries are
changed.
NOTE
Your meter uses two CR2032 3V lithium batteries. This type of batteries is
available in many stores. Keep spare batteries handy at all times.
You do not need to set the date and time if you insert new batteries within 1
minute after removing the old batteries. Even if you change the batteries within
1 minute, your meter’s clock may stop if:
The batteries are not replaced soon after the battery symbol appears.
The batteries are inserted upside down.
You touch the metal parts inside the meter with hands or metal.
1. Make sure the meter is off.
2. Grip the strap hole with the tip of your finger and pull
open.
3. Remove the batteries.
4. Set the new batteries into the battery holder with the
“+” side facing upward.
NOTE
If the batteries are inserted upside down, the meter will
not operate.
5. Close the battery holder.
6. The “SET” screen appears when the time has been reset. If it appears, set
the correct time. See the setup mode flowchart in chapter 7. If you do not set
the time, all subsequent test results will be stored as 12:00 AM or 00:00.
Dispose of old batteries according to your local environmental
regulations.
To obtain accurate test results:
Do not use test strips or control solutions if their bottles are broken or
have been left open.
Do not use the test strips or control solutions beyond their expiration
dates.
CAUTION
Comentarios a estos manuales